FRDD
Forensically-Relevant Drug Database
Description
Preferred Drug Name
Brief Description
THCP is a semi-synthetic cannabinoid categorized as a phytocannabinoid and is structurally similar to other phytocannabinoids (e.g., THC, CBD, CBDP). THCP is present at trace amounts in Cannabis sativa; however, drug products purportedly containing synthetically derived THCP have been sold online and in smoke shops in the form of plant materials, edibles, and vapes. THCP was first identified by our laboratory in May 2025 and confirmed after acquiring standard reference material.
Structure

Nomenclature
Formal Name (IUPAC)
Synonyms
Identifiers
CAS Numbers
Chemistry
Chemical Formula
Molecular Mass
Molecular Ion
Exact Mass
SMILES
InChI String
InChI Key
GCMS Analytics
Lab
Comments
| Sample Preparation: | Drug Materials: Dilution in methanol or acid / base extraction Drug Standards: Dilution in methanol |
| Instrument: | Agilent 5975 Series GC/MSD System |
| Column: | Agilent J&W DB-1 (12 m x 200 µm x 0.33 µm) |
| Carrier Gas: | Helium (Flow: 1.46 mL/min) |
| Temperatures: | Injection Port: 265 °C Transfer Line: 300 °C MS Source: 230 °C MS Quad: 150 °C Oven Program: 50 °C for 0 min, 30 °C/min to 340 °C for 2.3 min |
| Injection Parameters: | Injection Type: Splitless, Injection Volume: 1 µL |
| MS Parameters: | Mass Scan Range: 40-550 m/z |
Images

Fragments
342
259
299
243
HRMS Analytics
Lab
Comments
| Sample Preparation: | Toxicology Samples: Liquid-liquid extraction Drug Materials: Dilution of sample in mobile phase |
| Instrument: | Sciex TripleTOF® 5600+, Shimadzu Nexera XR UHPLC Sciex X500R, Sciex ExionLC |
| Column: | Phenomenex® Kinetex C18 (50 mm x 3.0 mm, 2.6 µm) |
| Mobile Phase: | A: Ammonium formate (10 mM, pH 3.0) B: Methanol/acetonitrile (50:50) Flow rate: 0.4 mL/min |
| Gradient: | Initial: 95A:5B; 5A:95B over 13 min; 95A:5B at 15.5 min |
| Temperatures: | Autosampler: 15 °C Column Oven: 30 °C Source Heater: 600 °C |
| Injection Parameters: | Injection Volume: 10 µL |
| QTOF Parameters: | TOF MS Scan Range: 100-510 Da Precursor Isolation: SWATH® acquisition (27 windows) Fragmentation: Collison Energy Spread (35±15 eV) MS/MS Scan Range: 50-510 Da |
Images


Fragments
221.1526
287.1994
135.1167
123.0443